Fault phenomenon:
There is no echo when using all the probes.
Failure analysis:
When using all the probes are no echo, consider it is RFM423 board or power supply failure.
Repair results:
At first, replacing the RFM423 board, the echo appeared and worked normally when connected with all the probes.